Reliability of Peak Exercise Testing in Patients with Heart Failure with Preserved Ejection Fraction

Jessica M Scott a, Mark J Haykowsky b, Joel Eggebeen c, Timothy M Morgan d, Peter H Brubaker e, Dalane W Kitzman c
PMCID: PMC3511643  NIHMSID: NIHMS404386  PMID: 22981266

Abstract

Exercise intolerance is the primary symptom among heart failure patients with preserved ejection fraction (HFpEF), is a major determinant of their reduced quality of life, and an important outcome in clinical trials. Although cardiopulmonary exercise testing (CPET) provides peak and submaximal diagnostic indices, the reliability of peak treadmill CPET in patients ≥ 55 years of age with HFpEF has not been examined. Two CPETs were performed in 52 HFpEF patients (age 70 ± 7 years). The two tests were separated by an average of 23 ± 13 days (median: 22 days) and performed under identical conditions, with no intervention or change in status between visits except for initiation of a placebo run-in. A multi-step protocol for patient screening, education, and quality control was utilized. Mean peak VO2 was similar on test 1 and test 2 (14.4 ± 2.4 vs. 14.3 ± 2.3 ml/kg/min). The correlation coefficients and intraclass correlations (ICC) from the testing days were as follows: VO2 r = 0.85, p < 0.001, ICC = 0.855; ventilatory anaerobic threshold r= 0.79, p < 0.001, ICC= 0.790; VE/VCO2 slope r = 0.87, p < 0.001, ICC = 0.864; HR r = 0.94, p < 0.001, ICC = 0.938. These results challenge conventional wisdom that serial baseline testing is required in clinical trials with exercise capacity outcomes. In conclusion, in women and men with HFpEF and severe physical dysfunction, key submaximal and peak exercise testing variables exhibited good reliability and were not significantly altered by a learning effect or placebo administration.

Keywords: heart failure preserved ejection fraction, cardiopulmonary exercise testing, test reproducibility

INTRODUCTION

Non-invasive assessment of exercise intolerance using cardiopulmonary exercise testing (CPET) provides crucial diagnostic data for the evaluation and treatment of patients with heart failure (HF), and is commonly used as a clinical endpoint in HF therapeutic trials.13 CPET guidelines have been developed and regularly updated by expert working groups in order to standardize both testing procedures and interpretation of exercise data.4,5 These guidelines commonly recommend that at least two CPETs be conducted: the first for habituation/learning purposes; the second for acquisition of baseline data.5,6 The stated purpose of this multiple test approach is to reduce test variability. However, this adds significantly to participant burden, and study time and cost. Furthermore, although the reliability of CPET has been examined in symptomatic HF and reduced EF (HFrEF),4 there are relatively few data verifying the reliability of peak exercise tests in patients ≥55 years, women, and patients with severely reduced exercise function who might benefit most by a habituation session. The latter patients are typical for the entity of HF with preserved EF (HFpEF), in which very few reliability data have been reported.7 Importantly, the impact of a placebo run-in period on CPET variables in this population, relevant to pharmacological interventions, has not previously been investigated. The purpose of this study was to determine the reliability of a treadmill exercise test with a placebo run-in period for assessment of peak and submaximal indices of exercise tolerance in women and men ≥55 years with HFpEF utilizing participant screening, well-trained testing personnel, and adherence to stringent CPET criteria.

METHODS

The study protocol was approved by the Wake Forest University Medical Center Institutional Review Board, and written informed consent was obtained from all participants. This report is from the screening visit (test 1) and baseline visit (test 2) of a randomized, double-blind, placebo controlled pharmacological clinical trial whose primary outcome is exercise capacity. Participant eligibility was assessed at an initial screening visit, which included medical history, physical examination, blood labs, echocardiogram to document ejection fraction and evidence of diastolic dysfunction, and peak CPET (test 1). Approximately one week later, participants returned for a second screening visit. If the initial screening visit blood labs were within inclusion criteria, participants entered into a two week placebo run-in period. At the subsequent baseline visit, if compliance to the placebo run-in by pill count was at least 80%, participants completed all baseline study assessments, including a second peak CPET (test 2). Patients had NYHA class II–III of at least 1 month duration with preserved resting ejection fraction, echocardiographic evidence of diastolic dysfunction (early diastolic lateral mitral annular velocity ≥9 cm/sec.), and no evidence of significant anemia or coronary artery, valvular, infiltrative, pericardial, pulmonary, or renal disease. Additionally, only participants with evidence of severe exercise intolerance (exercise time of at least 2 minutes but not greater than 12 minutes for women or 14 minutes for men) were enrolled. Potential participants were initially screened for inclusion by retrospective review of clinic visits and hospital discharge records at the Wake Forest University Medical Center. Of the resultant 273 preliminarily eligible participants, 66 were scheduled for an initial screening visit, and 52 met the study criteria and were enrolled.

The multi-step screening, education, and quality control CPET protocol is outlined in Appendix 1. Briefly, exercise testing was performed on a treadmill using the modified Naughton protocol and conducted by the same trained, experienced individual with master's degree in exercise physiology. Prior to each test, participants were screened for any recent acute illness or injury or change in status and were given detailed, standardized, verbal instructions. During the test, participants were encouraged to walk for as long as possible and, once fatigued, to attempt the subsequent stage, resulting in a peak, symptom-limited exhaustive effort. Complete calibration of gas concentrations and flow were performed before each test. Metabolic gas exchange was measured continuously during exercise and averaged over 15-second intervals (Medgraphics Ultima, Medical Graphics Corp., St. Paul, Minnesota). Peak VO2 was defined as the average of the 2 highest VO2 values for a given 15-second interval within the last 90 seconds of exercise. Ventilatory anaerobic threshold (VAT) was calculated using standard methods.8 The VE/VCO2 slope was determined by measuring the slope across the entire course of exercise.9 Oxygen uptake efficiency slope (OUES) was determined measuring the slope of VO2 (ml/min) and log10VE (L/min) across the entire course of exercise.10 All tests were evaluated by the exercise physiologist for validity based on a combination of symptoms (dyspnea on exertion and/or fatigue as primary endpoints) and physiological measures (respiratory exchange ratio, heart rate, minute ventilation, respiratory rate) and confirmed if necessary by a cardiologist. To avoid bias, the tester intentionally did not view the results of test 1 immediately before conducting test 2. Additionally, validation of the gas exchange unit was conducted prior to each testing day by performing a 2-stage, incremental exercise test in a known, constant test subject. In order to be validated, VO2 and VCO2 needed to be within 2 standard deviations of the usual, known value at each workload.

Means are expressed using either ± standard deviation or with 10th and 90th percentiles. Within-subject variability from test 1 to test 2 was quantified by the within-subject absolute change (i.e., either increase or decrease from test 1 to test 2). A paired t-test was used to test for a statistical difference between the test 1 and test 2 values. Bland-Altman plot was used to visually assess whether the magnitude of within-subject variability in peak VO2 varied with the magnitude of the measurements.11 Reliability of the CPET outcome measures between the screening visit and baseline visit was tested by Pearson correlations and intraclass correlations. One-sided 95% confidence intervals of Pearson correlations were made using Fisher's z transformation. For each variable, the coefficient of variation (CV) was defined as the (within-subject standard deviation/within-subject mean) × 100%. Significance was set at p<0.05.

RESULTS

Participants were predominantly white women with a history of hypertension and NYHA functional class II symptoms (Table 1). Test 1 and test 2 were separated by an average of 23 ± 13 days. Compliance by pill count to the placebo run-in period was excellent (96%). No adverse events occurred during or after any peak CPET. In three participants, the screening visit CPET was not considered to be a valid peak test and was repeated two weeks later. The reasons were: 1) abnormal gait and coordination resulting in early fatigue and test termination in two participants, and 2) test ended primarily due to knee pain in one participant. The two participants with abnormal gait and coordination were coached on proper mechanics immediately following the screening visit CPET with good results. The participant with initial knee pain returned following a bout of NSAIDS and completed the repeated CPET with no pain. In each case, the repeated screening visit CPET was considered to be valid and was used in this analysis.

Table 1.

Participant characteristics (n=52)

Age (years) 70 ± 7
Men 10 (19%)
Women 42 (81%)
White 31 (60%)
Black 21 (40%)
Weight (kg) 89 ± 17
Systolic blood pressure (mm Hg) 129 ± 18
Diastolic blood pressure (mm Hg) 74 ± 10
Body mass index (kg/m ) 33.5 ± 5.7
New York Heart Association functional class
 II 37 (71%)
 III 15 (29%)
Doppler diastolic function
 Abnormal relaxation 38 (73%)
 Pseudonormal 11 (21%)
 Undetermined (atrial fibrillation) 3 (6%)
Diabetes Mellitus 21 (40%)
Hypertension (by history) 50 (96%)
Medications
 angiotensin-converting enzyme inhibitors 20 (38%)
 Digoxin 1 (2%)
 Diuretics 37 (71%)
 Beta-blockers 24 (46%)
 Calcium-channel blockers 19 (37%)
Peak oxygen consumption, test 1 (ml/kg/min) 14.4 ± 2.4

Values are expressed as n (%) or mean ± SD.

Peak VO2 and VAT were nearly identical on the 2 tests. Within-subject peak VO2 demonstrated high agreement between tests as shown by Bland-Altman analysis (Figure 1). Reliability of CPET parameters was excellent (Table 2), and was similar across age and gender (Tables 3 and 4). Exercise time increased significantly from test 1 to test 2. Likewise, VE/VCO2 slope was significantly greater in test 2 than test 1 (Table 5).

Figure 1.

Figure 1

Bland-Altman plot for change in peak VO2 versus mean peak VO2 for test 1 and test 2.

Table 2.

Reliability of cardiopulmonary exercise testing

Test 1 Test 2 R 1-sided 95% CI ICC P-Value
Peak Exercise (treadmill)
Time (seconds) 577 ± 111 598 ± 123 0.914 0.866 0.895 <0.001
Indexed peak oxygen uptake (ml/kg/min) 14.4 ± 2.4 14.3 ± 2.3 0.853 0.775 0.855 <0.001
Absolute peak oxygen uptake (ml/min) 1266 ± 300 1260 ± 290 0.918 0.872 0.919 <0.001
Heart rate (beats/min) 128 ± 25 127 ± 24 0.938 0.903 0.938 <0.001
Systolic blood pressure (mmHg) 170 ± 18 167 ± 19 0.772 0.659 0.763 <0.001
Diastolic blood pressure (mmHg) 74 ± 10 72 ± 8 0.746 0.622 0.726 <0.001
Peak carbon dioxide production (ml/min) 1427 ± 346 1431 ± 346 0.893 0.834 0.895 <0.001
Respiratory exchange ratio 1.13 ± 0.06 1.14 ± 0.07 0.623 0.458 0.622 <0.001
Ventilation per carbon dioxide slope 30.5 ± 4.5 31.1 ± 4.5 0.872 0.802 0.864 <0.001
Oxygen uptake at ventilatory threshold (ml/min) 845 ± 213 840 ± 212 0.787 0.678 0.790 <0.001
Oxygen uptake efficiency slope (ml/min*L/min) 1567 ± 449 1553 ± 450 0.869 0.798 0.871 <0.001

P-Value corresponds to the R value.

Table 3.

Effects of age on cardiopulmonary exercise reliability

Age < 70 (n=28) Age ≥ 70 (n=24)
Test 1 Test 2 R ICC Test 1 Test 2 R ICC
Exercise time (sec) 593 ± 99 617 ± 100 0.90 0.878 557 ± 124 577 ± 145 0.92 0.904
Oxygen uptake (ml/kg/min) 14.2 ± 2.5 14.2 ± 2.5 0.89 0.894 14.6 ± 2.3 14.5 ± 2.3 0.80 0.806
Oxygen uptake (ml/min) 1305 ± 262 1295 ± 228 0.91 0.900 1220 ± 339 1220 ± 349 0.93 0.931
Heart rate (beats/min) 132 ± 25 132 ± 22 0.93 0.920 124 ± 26 121 ± 26 0.96 0.954
Systolic blood pressure (mmHg) 172 ± 17 169 ± 17 0.80 0.798 167 ± 19 164 ± 22 0.74 0.734
Diastolic blood pressure (mmHg) 76 ± 9 74 ± 8 0.77 0.742 72 ± 11 70 ± 8 0.71 0.697
Carbon dioxide production (ml/min) 1448 ± 286 1462 ± 257 0.87 0.867 1401 ± 410 1395 ± 430 0.91 0.910
Respiratory exchange ratio 1.12 ± 0.06 1.13 ± 0.06 0.68 0.661 1.15 ± 0.06 1.14 ± 0.08 0.60 0.578
Ventilation per carbon dioxide slope 29.7 ± 4.7 30.5 ± 4.9 0.88 0.875 31.3 ± 4.2 31.8 ± 4.0 0.85 0.843
Oxygen uptake at ventilatory threshold (ml/min) 861 ± 190 860 ± 163 0.76 0.755 827 ± 239 818 ± 258 0.81 0.814
Oxygen uptake efficiency slope 1642 ± 340 1641 ± 403 0.77 0.762 1480 ± 545 1451 ± 487 0.94 0.932

Table 4.

Effects of gender on cardiopulmonary exercise reliability

Women (n=42) Men (N=10)
Test 1 Test 2 R ICC Test 1 Test 2 R ICC
Exercise time (sec) 568 ± 107 590 ± 116 0.92 0.902 611 ± 126 632 ± 151 0.89 0.877
Oxygen uptake (ml/kg/min) 13.8 ± 1.9 13.9 ± 1.8 0.79 0.795 16.8 ± 2.7 16.3 ± 3.2 0.88 0.865
Oxygen uptake (ml/min) 1183 ± 234 1187 ± 221 0.89 0.888 1614 ± 306 1571 ± 348 0.89 0.885
Heart rate (beats/min) 133 ± 24 132 ± 21 0.93 0.927 110 ± 24 107 ± 26 0.94 0.934
Systolic blood pressure (mmHg) 171 ± 18 168 ± 20 0.80 0.789 165 ± 18 161 ± 16 0.61 0.616
Diastolic blood pressure (mmHg) 74 ± 9 72 ± 8 0.71 0.694 74 ± 12 72 ± 9 0.86 0.840
Carbon dioxide production (ml/min) 1321 ± 251 1338 ± 256 0.84 0.842 1870 ± 348 1824 ± 406 0.84 0.840
Respiratory exchange ratio 1.12 ± 0.06 1.13 ± 0.07 0.55 0.550 1.16 ± 0.06 1.17 ± 0.07 0.84 0.852
Ventilation per carbon dioxide slope 30.7 ± 4.5 31.5 ± 4.6 0.86 0.852 29.4 ± 4.9 29.7 ± 4.4 0.91 0.908
Oxygen uptake at ventilatory threshold (ml/min) 790 ± 170 783 ± 157 0.71 0.716 1070 ± 231 1074 ± 252 0.68 0.705
Oxygen uptake efficiency slope 1455 ± 333 1460 ± 378 0.81 0.811 2040 ± 577 1944 ± 532 0.89 0.886

Table 5.

Test-retest variability of peak exercise variables

Test 1 Test 2 P-Value Absolute Change (2 – 1) % 2 > 1 % 1 > 2 cv
Exercise Time (seconds) 577 (411, 720) 598 (439, 720) 0.003* 22 (−50, 90) 54 25 4.9%
Indexed oxygen uptake (ml/kg/min) 14.4 (11.1, 18.0) 14.3 (11.6, 18.0) 0.79 −0.05 (−1.50, 1.37) 40 52 4.9%
Ventilatory threshold (ml/min) 845 (581, 1177) 840 (603, 1064) 0.81 −5 (−175, 190) 45 53 9.3%
Ventilation per carbon dioxide slope 30.5 (25.3, 36.3) 31.1 (24.9, 36.8) 0.04* 0.7 (−2.2, 3.8) 65 35 4.5%
Respiratory exchange ratio 1.13 (1.07, 1.21) 1.14 (1.06, 1.24) 0.55 0.01 (−0.07, 0.07) 50 44 3.0%
Oxygen uptake efficiency slope (ml/min*L/min) 1567 (1074, 2071) 1553 (1107, 2076) 0.67 −14 (−243, 341) 40 60 8.5%

Values expressed as mean (10th percentile, 90th percentile) unless noted otherwise. P-Value corresponds to a paired t-test of the null hypothesis that the mean of test 2 minus the mean of test 1 = 0.

*

paired t-test of the null hypothesis that the mean of test 2 minus the mean of test 1 =0.

DISCUSSION

Few data are available regarding the reliability of CPET in women, and specifically in patients ≥55 years of age with HFpEF despite the fact that HFpEF occurs in approximately 50% or more of HF patients in the community, and the proportion is higher among women and individuals >55 years.1214 Since HFpEF patients are predominantly ≥55 years and female, they may be expected to be less familiar with exercise testing, to have greater variability, and to have a greater need for “learning” or practice sessions. Moreover, the impact of a placebo run-in period on CPET variables in this population, relevant to pharmacological interventions, has not previously been investigated. Accordingly, in this study we investigated the reliability of CPET-derived variables specifically in 52 patients ≥55 years with HFpEF who underwent 2 identical CPETs approximately 3 weeks apart with a placebo run-in period between tests. We utilized a multi-step screening, education, and quality control protocol to optimize reliability. The major new findings of this study were 1) No change in the mean value or significant within-subject variability in peak VO2 or VAT between tests, suggesting no significant learning or placebo effect, 2) Good correlations in all CPET parameters between tests, with no difference across age and gender. Thus, in patients with HFpEF ≥55 years of age, there may be no need to perform more than one CPET to obtain valid, baseline peak VO2 and other CPET results.

Numerous studies have shown that peak VO2 is a key outcome for assessing the prognosis of patients with HF, and is commonly used as a clinical endpoint in HF therapeutic trials, including those with HFpEF.1,2,15 Current cardiology guidelines recommend that at least two CPETs be conducted based on prior work demonstrating that peak VO2 can vary significantly over serial tests.5 For example, Elborn et al.16 performed three consecutive treadmill tests separated by two weeks on 30 subjects with HFrEF. The mean peak VO2 improved significantly from the first to the second test (14.1 vs. 14.9 mL/kg/min, p<0.005), while the average within-subject CV for the three tests was 6%. The authors concluded that a single baseline test was insufficient and suggested that the performance of at least two tests be conducted for clinical research applications. In contrast, Bensimhon et al.4 performed two baseline peak treadmill tests on 398 men and women with symptomatic HFrEF and found that there was no significant change in peak VO2 (15.2 ± 5.0 vs. 15.2 ± 5.0 mL/kg/min; p=0.78). Although within-subject variability of peak VO2 was significant, the authors concluded that in multi-center trials, there is no need to perform more than 1 baseline CPET. In the only study to examine HFpEF to date, Marburger et al.7 examined the reproducibility of a cycle ergometer CPET in 5 HFpEF patients and 4 HFrEF patients. The authors concluded that the reliability of peak testing was good in both HFpEF and HFrEF, and that only one test was necessary. However, the small number of patients with HFpEF significantly limits the widespread applicability of results. In the present study with 52 HFpEF patients, for peak VO2, there was good correlation between tests (r = 0.85; ICC = 0.855), the overall group means on test 1 and test 2 were essentially identical, the percentage of subjects who decreased or increased on test 2 were similar, Bland-Altman plots demonstrated high agreement between the 2 tests, and the within-subject variability was low (average CV <5%). Taken together, these findings further suggest that a single baseline test is sufficient to measure peak VO2 for the evaluation of therapy or assessment of prognosis.

A large body of research has also evolved regarding CPET variables other than peak VO2. Responses such as VAT, VE/VCO2 slope, OUES, and exercise time have been used with greater frequency to classify functional limitations and to stratify risk in patients with HF.5 VAT and OUES are independent of effort and protocol and, similar to peak VO2, have been shown to be highly reproducible.4,10 In this study, we demonstrate that VAT and OUES are also reproducible with repeat testing as OUES and VAT had good correlation between tests and nearly identical mean values on test 1 and 2. As seen previously in HFrEF,4,16,17 we found that the mean exercise time increased from test 1 to test 2, with 54% of subjects increasing on test 2. Additionally, mean values for VE/VCO2 slope increased from test 1 to test 2, and there were a greater number of subjects who increased on test 2 than decreased. The reasons for discordant findings for VE/VCO2 slope are uncertain; however, may be related to daily variations in ventilatory oscillation. Guazzi et al.18 reported that the prevalence of abnormal ventilatory oscillation in patients with HFpEF approximates 30%, which could affect VE/CO2.

To achieve these results and be able to conduct a single, valid baseline CPET in longitudinal studies, there are several important aspects to consider (see appendix). These include: standardized participant instructions, screening for intercurrent acute illness or injury or change in condition, frequent calibration and validation of the gas exchange unit, strong encouragement to reach a peak level, and evaluation of reasons for termination and physiological responses. CPET that do not meet these criteria will likely have reduced reliability. Although inter-observer variability was not assessed in the present study given that each test was conducted by the same experienced individual, high reproducibility is obtainable with appropriate training of laboratory personnel.4,7 These results challenge conventional wisdom that serial baseline testing is required in clinical trials with exercise capacity outcomes, and suggest that any incremental gain be balanced with the increase in participant burden and cost.

Supplementary Material

01

Acknowledgments

Funding: This study was supported in part by N.I.H. grants R37AG18915 and P30AG21332 and by an investigator-initiated grant from Novartis Pharmaceuticals.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Conflicts of Interest: Dr. Kitzman is a consultant for Relypsa Inc., Boston Scientific Corp., and Abbot, has received grant support from Novartis, and owns stock in Gilead Sciences. Dr. Brubaker is a consultant to and has received research support from Boston Scientific Corp. All other authors report no relationships to disclose.

References

  • 1.Haykowsky MJ, Brubaker PH, John JM, Stewart KP, Morgan TM, Kitzman DW. Determinants of exercise intolerance in elderly heart failure patients with preserved ejection fraction. J Am Coll Cardiol. 2012;58:265–274. doi: 10.1016/j.jacc.2011.02.055. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kitzman DW, Daniel KR. Diastolic heart failure in the elderly. Clin Geriatr Med. 2007;23:83–106. doi: 10.1016/j.cger.2006.09.002. [DOI] [PubMed] [Google Scholar]
  • 3.Haykowsky M, Brubaker P, Kitzman D. Role of physical training in heart failure with preserved ejection fraction. Curr Heart Fail Rep. 2012;9:101–106. doi: 10.1007/s11897-012-0087-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bensimhon DR, Leifer ES, Ellis SJ, Fleg JL, Keteyian SJ, Pina IL, Kitzman DW, McKelvie RS, Kraus WE, Forman DE, Kao AJ, Whellan DJ, O'Connor CM, Russell SD. Reproducibility of peak oxygen uptake and other cardiopulmonary exercise testing parameters in patients with heart failure (from the Heart Failure and A Controlled Trial Investigating Outcomes of exercise traiNing) Am J Cardiol. 2008;102:712–717. doi: 10.1016/j.amjcard.2008.04.047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Arena R, Myers J, Williams MA, Gulati M, Kligfield P, Balady GJ, Collins E, Fletcher G. Assessment of functional capacity in clinical and research settings: a scientific statement from the American Heart Association Committee on Exercise, Rehabilitation, and Prevention of the Council on Clinical Cardiology and the Council on Cardiovascular Nursing. Circulation. 2007;116:329–343. doi: 10.1161/CIRCULATIONAHA.106.184461. [DOI] [PubMed] [Google Scholar]
  • 6.Myers J, Froelicher VF. Optimizing the exercise test for pharmacological investigations. Circulation. 1990;82:1839–1846. doi: 10.1161/01.cir.82.5.1839. [DOI] [PubMed] [Google Scholar]
  • 7.Marburger CT, Brubaker PH, Pollock WE, Morgan TM, Kitzman DW. Reproducibility of cardiopulmonary exercise testing in elderly patients with congestive heart failure. Am J Cardiol. 1998;82:905–909. doi: 10.1016/s0002-9149(98)00502-5. [DOI] [PubMed] [Google Scholar]
  • 8.Beaver WL, Wasserman K, Whipp BJ. A new method for detecting anaerobic threshold by gas exchange. J Appl Physiol. 1986;60:2020–2027. doi: 10.1152/jappl.1986.60.6.2020. [DOI] [PubMed] [Google Scholar]
  • 9.Arena R, Myers J, Aslam SS, Varughese EB, Peberdy MA. Technical considerations related to the minute ventilation/carbon dioxide output slope in patients with heart failure. Chest. 2003;124:720–727. doi: 10.1378/chest.124.2.720. [DOI] [PubMed] [Google Scholar]
  • 10.Baba R, Nagashima M, Goto M, Nagano Y, Yokota M, Tauchi N, Nishibata K. Oxygen uptake efficiency slope: a new index of cardiorespiratory functional reserve derived from the relation between oxygen uptake and minute ventilation during incremental exercise. J Am Coll Cardiol. 1996;28:1567–1572. doi: 10.1016/s0735-1097(96)00412-3. [DOI] [PubMed] [Google Scholar]
  • 11.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310. [PubMed] [Google Scholar]
  • 12.Kitzman DW, Gardin JM, Gottdiener JS, Arnold A, Boineau R, Aurigemma G, Marino EK, Lyles M, Cushman M, Enright PL. Importance of heart failure with preserved systolic function in patients > or = 65 years of age. CHS Research Group. Cardiovascular Health Study. Am J Cardiol. 2001;87:413–419. doi: 10.1016/s0002-9149(00)01393-x. [DOI] [PubMed] [Google Scholar]
  • 13.Vasan RS, Larson MG, Benjamin EJ, Evans JC, Reiss CK, Levy D. Congestive heart failure in subjects with normal versus reduced left ventricular ejection fraction: prevalence and mortality in a population-based cohort. J Am Coll Cardiol. 1999;33:1948–1955. doi: 10.1016/s0735-1097(99)00118-7. [DOI] [PubMed] [Google Scholar]
  • 14.Bhatia RS, Tu JV, Lee DS, Austin PC, Fang J, Haouzi A, Gong Y, Liu PP. Outcome of heart failure with preserved ejection fraction in a population-based study. N Engl J Med. 2006;355:260–269. doi: 10.1056/NEJMoa051530. [DOI] [PubMed] [Google Scholar]
  • 15.Haykowsky M, Brubaker P, Kitzman D. Role of Physical Training in Heart Failure with Preserved Ejection Fraction. Curr Heart Fail Rep. 2011:2011. doi: 10.1007/s11897-012-0087-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Elborn JS, Stanford CF, Nicholls DP. Reproducibility of cardiopulmonary parameters during exercise in patients with chronic cardiac failure. The need for a preliminary test. Eur Heart J. 1990;11:75–81. doi: 10.1093/oxfordjournals.eurheartj.a059596. [DOI] [PubMed] [Google Scholar]
  • 17.Sullivan M, Genter F, Savvides M, Roberts M, Myers J, Froelicher V. The reproducibility of hemodynamic, electrocardiographic, and gas exchange data during treadmill exercise in patients with stable angina pectoris. Chest. 1984;86:375–382. doi: 10.1378/chest.86.3.375. [DOI] [PubMed] [Google Scholar]
  • 18.Guazzi M, Myers J, Peberdy MA, Bensimhon D, Chase P, Arena R. Exercise oscillatory breathing in diastolic heart failure: prevalence and prognostic insights. Eur Heart J. 2008;29:2751–2759. doi: 10.1093/eurheartj/ehn437. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

01

RESOURCES