. 2020 Feb 5;9(2):370. doi: 10.3390/cells9020370

Table 3.

Application of all three methods on independent dataset. Fifteen different CSF samples were analyzed with hemoglobin ELISA, Combur10-Test Strip, and LC-MS. The respective results were classified to the defined contamination levels as listed in Table 2. Samples that were assigned by all three approaches to the same blood level are highlighted in light gray, and samples that were assigned to the same level by two approaches are marked in dark gray.

Sample ELISA strip LC-MS
CSF 5 1 1 1
CSF 6 1 1 1
CSF 7 10 1 10
CSF 8 1 1 1
CSF 9 10 1 1
CSF 10 10 10 10
CSF 11 20 20 20
CSF 12 20 20 20
CSF 13 20 10 10
CSF 14 10 10 10
CSF 15 20 40 40
CSF 16 20 20 20
CSF 17 20 20 20
CSF 18 30 40 40
CSF 19 20 40 30