Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2014 Jul 24.
Published in final edited form as: J Vis Exp. 2014 May 27;(87):10.3791/51358. doi: 10.3791/51358

A strategy for sensitive, large scale quantitative metabolomics

Xiaojing Liu 1, Zheng Ser 1, Ahmad A Cluntun 1,2, Samantha J Mentch 1,2, Jason W Locasale 1,2
PMCID: PMC4109060  NIHMSID: NIHMS602735  PMID: 24894601

Abstract

Short Abstract

Metabolite profiling has been a valuable asset in the study of metabolism in health and disease. Utilizing normal-phased liquid chromatography coupled to high-resolution mass spectrometry with polarity switching and a rapid duty cycle, we describe a protocol to analyze the polar metabolic composition of biological material with high sensitivity, accuracy and resolution.

Long Abstract

Metabolite profiling has been a valuable asset in the study of metabolism in health and disease. However, current platforms have different limiting factors, such as labor intensive sample preparations, low detection limits, slow scan speeds, intensive method optimization for each metabolite, and the inability to measure both positively and negatively charged ions in single experiments. Therefore, a novel metabolomics protocol could advance metabolomics studies. Amide-based hydrophilic chromatography enables polar metabolite analysis without any chemical derivatization. High resolution MS using the Q-Exactive (QE-MS) has improved ion optics, increased scan speeds (256 ms at resolution 70 000), and has the capability of carrying out positive/negative switching. Using a cold methanol extraction strategy, and coupling an amide column with QE-MS enables robust detection of 168 targeted polar metabolites and thousands of additional features simultaneously. Data processing is carried out with commercially available software in a highly efficient way, and unknown features extracted from the mass spectra can be queried in databases.

Keywords: high-resolution mass spectrometry, metabolomics, positive/negative switching, low mass calibration, Orbitrap

Introduction

Metabolomics, defined as an experiment that measures multiple metabolites simultaneously, has been an area of intense interest. Metabolomics provides a direct readout of molecular physiology and has provided insights into development and disease such as cancer14. Nuclear magnetic resonance (NMR) and Gas chromatography-mass spectrometry (GC-MS) are among the most commonly used instruments59. NMR, especially has been used for flux experiments since heavy isotope labeled compounds, such as 13C labeled metabolites, are NMR-active10,11. However, this strategy requires relatively high sample purity and large sample quantity, which limits its applications in metabolomics. Meanwhile, data collected from NMR needs intensive analysis and compound assignment of complex NMR spectra is difficult. GC-MS has been widely used for polar metabolites and lipid studies, but it requires volatile compounds and therefore often derivatization of metabolites, which sometimes involves complex chemistry that can be time consuming and introduces experimental noise.

Liquid Chromatography (LC) coupled to triple quadrupole Mass Spectrometry uses the first quadrupole for selecting the intact parent ions, which are then fragmented in the second quadrupole, while the third quadrupole is used to select characteristic fragments or daughter ions. This method, which records the transition from parent ions to specific daughter ions, is termed multiple reaction monitoring (MRM). MRM is a very sensitive, specific and robust method for both small molecule and protein quantitation1215, 21. However, MRM does have its limitations. To achieve high specificity a MRM method needs to be built for each metabolite. This method consists of identifying a specific fragment and corresponding optimized collision energy, which requires pre-knowledge of the properties of the metabolites of interest, such as chemical structure information. Therefore, with some exceptions involving the neutral loss of common fragments, it is not possible to identify unknown metabolites with this method.

In the recent years, high-resolution mass spectrometry (HRMS) instruments have been released, such as the LTQ-orbitrap and Exactive series, the QuanTof, and TripleTOF 5600.1618, 22 HRMS can provide a mass to charge ratio (m/z) of intact ions within an error of a few ppm. Therefore, an HRMS instrument operated by detecting all precursor ions (i.e. full scan mode) can obtain direct structural information from the exact mass and the resulting elemental composition of the analyte, and this information can be used to identify potential metabolites. Indeed, all information about a compound can be obtained with an exact mass, up to the level of structural isomers. Also, a full scan method does not require previous knowledge of metabolites and does not require method optimization. Moreover, since all ions with m/z falling into the scan range can be analyzed, HRMS has a nearly unlimited capacity in terms of the number of metabolites that can be quantified in a single run compared to the MRM method. HRMS is also comparable to a triple quadrupole MRM in quantitative capacity due to the short duty cycle resulting in a comparable number of data points that can be obtained in a full MS scan. Therefore, HRMS provides an alternative approach for quantitative metabolomics. Recently, an improved version of HRMS termed Q-Exactive mass spectrometry (QE-MS) can be operated under the switching between positive and negative modes with sufficiently fast cycle times in a single method, which expands the detection range19. Here we describe our metabolomics strategy using the QE-MS.

Protocol Text

1) Preparation of LC-MS reagents, establishment of a chromatography method, and establishment of instrument operating procedures

  • 1.1)

    Preparation of LC solvents.

  • 1.1.1)

    Prepare 500 ml mobile phase. A is 20 mM ammonium acetate and 15 mM ammonium hydroxide in 3% acetonitrile/water, final PH 9.0, and B is 100% acetonitrile.

  • 1.1.2)

    Loosely cap the bottle and place it in a water bath sonicator, and sonicate for 10 min without extra heating. (This step is to ensure that all of the ammonium salts completely dissolve and that there are no residual air bubbles.)

  • 1.1.3)

    Transfer 250 ml of the solvent to a 250 ml glass bottle for LC-MS use, and keep the remainder at 4 °C.

  • 1.2)

    Prepare the low mass range calibration solution. It is important to use a customized low mass range calibration mixture for metabolomics applications to ensure that accurate masses are detected at low molecular weights.

  • 1.2.1)

    Weigh 5 mg of both sodium fluoroacetate and homovanillic acid and dissolve them into 5 ml water to make a final concentration of 1 mg/ml. Dissolve diazinon in methanol to make a final concentration of 10 μg/ml.

  • 1.2.2)

    To prepare 1 ml of negative low mass calibration solution, mix 960 μl of thermo negative calibration solution with 20 μl of sodium fluoroaceate and homovanillic acid solution. To make 1 ml of positive low mass calibration solution, mix 990 μl of thermo positive calibration solution and 10 μl diazinon solution. (The low mass calibration solution should be stored at 4 °C and be prepared fresh every two months.)

  • 1.3)

    Calibration of QE-MS at a low mass range.

  • 1.3.1)

    Before performing low mass range calibration, carry out a standard mass calibration (m/z, 150–2000) in both positive and negative modes based on the manufacturer’s instructions.

  • 1.3.2)

    Once a regular mass calibration passes, adjust the scan range to 60–900 m/z in the instrument control panel and a source CID of 25 eV for positive mode and 35 eV for negative mode is applied. (This will give robust signals of the caffeine fragment ion and the sulfate ion. The scan range here is fixed, because the last m/z shouldn’t be larger than 15 times of the starting m/z)

  • 1.3.3)

    Input customized calibration ions (as listed in Table 1), and once the ion source is stable, then start the customized calibration in the computer tuning page. Note: A stable source is defined as less than 10 % of the total ion current variation in positive mode, and less than 15 % in negative mode.

  • 1.4)

    Establish the LC-MS instrumentation for polar metabolite analysis. LC is coupled to a QE-MS for metabolite separation and detection.

  • 1.4.1)

    Equip the QE-MS with a Heated electrospray ionization probe (H-ESI) by fixing it at level C and then connecting it to Nitrogen gas inlets, vaporizer cable and voltage cables. In the computer tuning page, set the relevant tuning parameters for the probe as listed: heater temperature, 120 °C; sheath gas, 30; auxiliary gas, 10; sweep gas, 3; spray voltage, 3.6 kV for positive mode and 2.5 kV for negative mode. Set the capillary temperature at 320 °C, and S-lens at 55.

  • 1.4.2)

    In the method program, build a full scan method as follows: Full scan range: 60 to 900 (m/z); resolution: 70, 000; maximum injection time: 200 ms with typical injection times around 50 ms; automatic gain control (AGC): 3,000,000 ions. These settings result in a duty cycle of around 550 ms to carry out scans in both positive and negative mode.

  • 1.4.3)

    Establish the chromatography method by inputting the linear gradient information. Employ an amide column (100 x 2.1 mm i.d., 3.5 μm) for compound separation at room temperature13,15.The mobile phase A is as described above, and mobile phase B is acetonitrile. Use a linear gradient as follows: 0 min, 85% B; 1.5 min, 85% B, 5.5 min, 35% B; 10min, 35% B, 10.5 min, 35% B, 14.5 min, 35% B, 15 min, 85% B, and 20 min, 85% B. The flow rate is 0.15 ml/min from 0 to 10 min and 15 to 20 min, and 0.3 ml/min from 10.5 to 14.5 min.

Table 1.

Low mass range calibration standards and their exact m/z.

The formula shown here is corresponding to the neutral form formula, and m/z is the neutral mass plus or minus a proton.

CSID Name Formula Monoisoto pic Mass Search Mass DeltaPPM R.T. (min)
234 beta-Alanine C3H7NO2 89.04800 89.04805 0.62 8.08
1057 Sarcosine C3H7NO2 89.04768 89.04805 4.22 8.08
5735 alanine C3H7NO2 89.04768 89.04805 4.22 8.08
568 Creatinine C4H7N3O 113.05900 113.05889 1.00 4.41
128566 Proline C5H9NO2 115.06333 115.06338 0.41 7.54
6050 L-(+)-Valine C5H11NO2 117.07898 117.07896 0.18 7.42
7762 Amyl nitrite I C5H11NO2 117.07898 117.07896 0.18 7.42
135 5-amino valeric acid C5H11NO2 117.07900 117.07896 0.38 7.42
242 trimethylglycine C5H11NO2 117.07900 117.07896 0.38 7.42
911 Niacinamide C6H6N2O 122.04800 122.04793 0.55 2.60
1091 Taurin C2H7NO3S 125.01466 125.01469 0.20 7.61
1030 Pyrroline hydroxycarboxylic acid C5H7NO3 129.04259 129.04259 0.02 8.51
7127 PCA C5H7NO3 129.04259 129.04259 0.02 8.51
90657 N-Acryloylglycine C5H7NO3 129.04259 129.04259 0.02 8.51
388752 5-Oxo-D-prolin C5H7NO3 129.04259 129.04259 0.02 8.51
389257 3-Hydroxy-3,4-dihydro-2H-pyrrole-5-carboxylic acid C5H7NO3 129.04259 129.04259 0.02 8.51
8031176 Pyrrolidonecarboxylic acid C5H7NO3 129.04259 129.04259 0.03 8.51
5605 Hydroxyproline C5H9NO3 131.05824 131.05901 5.83 8.08
7068 N-Acetylalanin C5H9NO3 131.05824 131.05901 5.83 8.08
79449 Ac-Ala-OH C5H9NO3 131.05824 131.05901 5.83 8.08
89122 Ethylformylglycine C5H9NO3 131.05824 131.05901 5.83 8.08
167744 l-Glutamic-gamma-semialdehyde C5H9NO3 131.05824 131.05901 5.83 8.08
388519 5-Amino-2-oxopentanoic acid C5H9NO3 131.05824 131.05901 5.83 8.08
134 Aminolevulinic acid C5H9NO3 131.05800 131.05901 7.70 8.08
9312313 3-Hydroxy-L-proline C5H9NO3 131.05800 131.05901 7.70 8.08
566 Creatine C4H9N3O2 131.06900 131.06905 0.36 8.08
5880 L-(+)-Leucine C6H13NO2 131.09464 131.09455 0.68 6.96
6067 L-(+)-Isoleucine C6H13NO2 131.09464 131.09455 0.68 6.96
19964 L-Norleucine C6H13NO2 131.09464 131.09455 0.68 6.96
388796 beta-Leucine C6H13NO2 131.09464 131.09455 0.68 6.96
548 Aminocaproic acid C6H13NO2 131.09500 131.09455 3.48 6.96
6031 L-(-)-Asparagine C4H8N2O3 132.05350 132.05348 0.10 8.31
109 Ureidopropionic acid C4H8N2O3 132.05299 132.05348 3.71 8.31
6026 L-Ornithine C5H12N2O2 132.08987 132.08988 0.02 10.37
64236 D-Ornithine C5H12N2O2 132.08987 132.08988 0.02 10.37
5746 Glutamine C5H10N2O3 146.06914 146.06900 0.93 8.25
128633 D-Glutamine C5H10N2O3 146.06914 146.06900 0.93 8.25
141172 Ureidoisobutyric acid C5H10N2O3 146.06914 146.06900 0.93 8.25
21436 N-Methyl-D-aspartic acid C5H9NO4 147.05316 147.05300 1.13 8.06
21814 D-(-)-Glutamic acid C5H9NO4 147.05316 147.05300 1.13 8.06
30572 L-(+)-Glutamic acid C5H9NO4 147.05316 147.05300 1.13 8.06
58744 N-Acetyl-L-serine C5H9NO4 147.05316 147.05300 1.13 8.06
5907 L-(-)-methionine C5H11NO2S 149.05106 149.05095 0.70 7.39
6038 Histidine C6H9N3O2 155.06947 155.06940 0.48 8.33
5910 L-(-)-Phenylalanine C9H11NO2 165.07898 165.07887 0.63 6.60
1025 Pyridoxine C8H11NO3 169.07390 169.07376 0.80 3.24
4463 Oxidopamine [USAN:INN] C8H11NO3 169.07390 169.07376 0.80 3.24
102750 5-(2-aminoethyl)-Pyrogallol C8H11NO3 169.07390 169.07376 0.80 3.24
388394 Norepinephrine C8H11NO3 169.07390 169.07376 0.80 3.24
6082 L-(+)-Arginine C6H14N4O2 174.11168 174.11144 1.39 10.77
64224 D-Arg C6H14N4O2 174.11168 174.11144 1.39 10.77
780 heteroauxin C10H9NO2 175.06300 175.06304 0.18 2.32
67261 Indole-3-acetaldehyde, 5-hydroxy- C10H9NO2 175.06332 175.06304 1.65 2.32
3574185 INDOLE-2-ACETIC ACID C10H9NO2 175.06332 175.06304 1.65 2.32
5833 L-(-)-Tyrosine C9H11NO3 181.07390 181.07378 0.66 7.48
389285 3-Amino-3-(4-hydroxyphenyl)propanoic acid C9H11NO3 181.07390 181.07378 0.66 7.48
13628311 L-threo-3-phenylserine C9H11NO3 181.07390 181.07378 0.66 7.48
425 4-hydroxy-4-(3-pyridyl)butanoic acid C9H11NO3 181.07401 181.07378 1.25 7.48
13899 3-(1H-Indol-3-yl)acrylic acid C11H9NO2 187.06332 187.06330 0.15 6.44
10607876 Indoleacrylic acid C11H9NO2 187.06332 187.06330 0.15 6.44
389120 N6,N6,N6-Trimethyl-L-lysine C9H20N2O2 188.15248 188.15221 1.45 10.87
388321 5"-S-Methyl-5"-thioadenosine C11H15N5O3S 297.08957 297.08898 2.00 2.56
144 9-(5-s-methyl-5-thiopentofuranosyl)-9h-purin-6-amine C11H15N5O3S 297.09000 297.08898 3.43 2.56
111188 Glutathione C10H17N3O6S 307.08380 307.08345 1.14 8.02

2) Preparation of metabolite samples

  • 2.1)

    Prepare an extraction solvent. Manually mix 40 ml methanol (LC-MS grade) and 10 ml water (LC-MS grade) in a 50 ml tube, and keep it in −80 °C freezer for at least 1 hour before use. NOTE: This procedure and the steps below can be modified for the extraction of biological tissue and fluid samples.

  • 2.2.1)

    Culture colon cancer HCT 8 cells in three 10 cm dishes or 6-well plates with full growth medium, RPMI 1640 supplemented with 10 % heat inactivated Fetal Bovine Serum and 100,000 units/L penicillin and 100 mg/L streptomycin.

  • 2.2.2)

    When cells reach 80 % confluence, quickly aspirate the medium, and place the dish or plate on top of dry ice13,15. Add 1ml extraction solvent immediately (80 % methanol/water), and transfer the plate to the −80 °C freezer. For 10 cm dish, 3 ml of extraction solvent is added to each well. (Try to remove the medium as much as possible to avoid the ion suppression effect due to residual salts from medium.)

  • 2.2.3)

    Leave the plate for 15 min. Remove it from the freezer, and scrape cells into the solvent on dry ice. Transfer the solution to 1.7 ml eppendorf tubes, and centrifuge with the speed of 20 000 × g at 4 °C for 10 min. (Prepare cell metabolites from three separate dishes to make three replicate samples. The purpose of keeping two tubes is to have one as a backup.)

  • 2.2.4)

    Transfer the supernatant to two new eppendorf tubes, and dry them in a speed vacuum. This takes about 3–6 hours depending on the speed vacuum used. (The samples can also be dried overnight under Nitrogen gas.)

  • 2.2.5)

    After drying, store tubes of each sample in the −80 °C freezer. When ready, take out samples and allow them to warm to ice temperature. Add 20 μl ice cold water (LC-MS grade), vortex to dissolve metabolites, and then centrifuge at 20 000 × g at 4 °C for 2 min. Transfer supernatant to LC vials, and inject 5 μl to LC-QE-MS for analysis.

3.) Setup of sample sequence

  • 3.1)

    Once the calibration has been properly carried out on the QE-MS, equilibrate LC column for five minutes with 85 % B at a flow rate of 0.15 ml/min, which is the starting condition of the LC gradient.

  • 3.2)

    Set up the sample sequence in random order. Note: In this way, it distributes the fluctuations introduced by the LC-MS to each sample and ensures more accurate comparison between different samples. Every 6 samples, add a wash run, which shares the same MS method, except the LC gradient is 95 % A for 10 min and followed by a 5 min column equilibration at 85 % A with a flow rate of 0.15 ml/min. Add a blank sample (100 % water) after each wash run to assess the system background and carry over levels.

  • 3.3)

    Save the sequence and once the LC column shows stable pressure, close to 400 psi, start the sequence run. If there is no other sample is to be run after this sequence, then add a stop run in the end of the sequence, which has a flow rate of 0 ml/min in the end of the gradient and choose “standby” after finishing the sequence.

  • 3.4)

    Re-run the same sample set 12 hrs after calibration. (This is to assess the mass error fluctuation after calibration.)

4.) Post analysis instrument cleaning and maintenance

  • 4.1)

    At the end of sequence, wash the column with 95 % A at a flow rate of 0.2 ml/min for 2 hrs, and if necessary, reverse the column before washing.

  • 4.2)

    Remove the LC column and directly connect the LC to the ion source by a union. Prepare cleaning solvent, water/methanol/formic acid (v:v:v, 90:10:0.2), set MS on standby mode, and wash LC-MS system at a flow rate of 0.1 ml/min for 1 hr to remove the residual precipitated salts or other impurities. Lower the flow rate if there is too much system pressure.

  • 4.3)

    Set the capillary temperature at 50 °C, and remove the ion cage. Carefully take out the ion sweep cone and the ion transfer tube after the capillary temperature drops to 50 °C. Use a rough mesh, such as sandpaper, to remove impurities left on the surface of the ion sweep cone.

  • 4.4)

    Place the ion transfer tube into a 15 ml falcon tube containing 10 ml 90 % water/methanol with 0.1 % formic acid. After sonicating the tube in a water bath sonicator for 20 min, decant the solvent inside, replace it with 10 ml pure methanol and sonicate for another 20 min. (If necessary, the sonication can be done at 40 °C or an even higher temperature to achieve better cleaning results.)

5.) Analysis of LC-MS data

  • 5.1)

    To ensure the sample sequence runs smoothly, after finishing the first two samples in the sequence, check peaks for unknown metabolites. Use a csv file listing metabolite names, neutral chemical formula and detection mode (either positive or negative), as the input file, and the output file contains extracted peaks and mass error in ppm. If the peak shape is abnormal or the mass error is off by more than 5 ppm, then the rest of the sequence needs to be stopped and troubleshooting needs to be done.

  • 5.2)

    After all of the samples in the sequence are finished, perform data analysis on a separate computer. Choose the method of “peak alignment and frame extraction” for small molecule on a commercial available software. load LC-MS raw data, and group them group them based on sample types. Pick samples in the middle of the run sequence as chromatography reference sample for peak alignment and pick a group as ratio group for fold change calculation. Use default frame parameter settings. Upload a frame seed including known metabolites for the targeted metabolites analysis with data collected in positive mode and a frame seed of metabolites for negative mode.

  • 5.3)

    Next, choose full spectra scan either in the positive or negative mode save this data processing file in the same folder as raw data. Turn off the database search function and run the workflow. Export the processed data as an excel sheet containing peak area of every frame. The first sets of frames correspond to the metabolites in the targeted list. Note: For a targeted metabolite analysis, the metabolites information is obtained based on the previous studies13,15.

  • 5.4)

    for an untargeted metabolite analysis, choose the method of “component extraction”. Load samples raw data and also three blank samples for background subtraction. Group raw data and set reference samples, as described previously. Use default frame parameters. Set component intensity threshold to 105, m/z width of 10 ppm and signal to noise ratio of 3.

  • 5.5)

    Use human metabolome database for unknown compounds identification. Use a coefficients of variation (CV) filter to remove the components with large CV within replicate samples. Manually go through each component and pick those with well-defined peak or relatively big difference in different samples types for database search. Export data with hits in database. (Peak alignment could be bypassed if the peak alignment score is too low.)

Representative Results

The accuracy of metabolomics data highly depends on the LC-QE-MS instrument performance. To assess whether the instrument is operating in good condition, and whether the method applied is proper, several known metabolite LC peaks are extracted from the total ion chromatography (TIC), as shown in Figure 1. Polar metabolites, including amino acids, glycolysis intermediates, TCA intermediates, nucleotides, vitamins, ATP, NADP+ and so on have good retention on the column and good peak shapes in the amide column under current LC conditions. Meanwhile, a mass error test is done within 24 hrs after low mass calibration, as illustrated in Figure 2. 6 different concentrations of samples in triplicate are run twice after calibration, and the whole time range covers almost 24 hrs. The mass error is assessed by comparing the detected m/z to the theoretical m/z of targeted metabolites. Here the targeted metabolites have an m/z ranging from 74 (glycine) to 744 (NADP+). The Y axis here represents the accumulative percentage of metabolites within certain mass error range. The red curve shows the result in positive mode, while the blue colored curve shows the data collected in negative mode. Figure 2 clearly indicates that more than 90% of metabolites are within 5 ppm mass error in both positive and negative mode, which means the low mass range calibration method developed here is sufficient to maintain 5 ppm mass error for low mass range detection.

Figure 1. Examples of LC-MS chromatography peaks.

Figure 1

Here, the reconstructed chromatography is generated with a mass window of 10 ppm (m/z ± 5 ppm). The X axis shows the retention time, while the Y axis shows the relative intensity, and the peak intensity is listed above every metabolite. Figure 1A shows peaks detected from positive mode, while figure 1B shows peaks from negative mode.

Figure 2. Evaluation of low mass range calibration.

Figure 2

The Y axis is the cumulative percentage of metabolites with mass detection error within 5 ppm. The X axis is the mass error range in ppm. Blue and red curves represent o to 12 hrs and 12 to 24 hrs, respectively.

Another issue to be addressed is the sensitivity of the instrument with the current method and instrument setup. A serial dilution of triplicate samples from 10 cm petri dish was done 5 times with a dilution factor of 6, ending up with 6 different concentrations of samples. These samples represent the amount of metabolites extracted from 107, 1.67×106, 2.78×105, 4.63×104, 7.72×103, and 1.29×103 of cells respectively. Since each concentration of sample is prepared in triplicate, a total of 18 samples are analyzed in LC-QE-MS. A targeted list is used to assess the number of metabolites detected at differing concentration of sample. The result in Figure 3 indicates that the optimal number of targeted metabolites detected is between 2.78×105 and 1.67×106 cells, while 1×107 cells give a fewer number of detected metabolites, which is due to ion suppression effects. This result indicates that the optimal amount of cells to extract for this analysis is roughly that of a well of in a 6-well plate.

Figure 3. Evaluation of sample amount - number of targeted metabolites detected versus number of HCT 8 cells.

Figure 3

Red squares represent metabolites detected in positive mode, blue circles mean metabolites measured in negative mode, and the black triangles are the total numbers of metabolites from both positive and negative mode. The X axis shows the number of HCT 8 cells.

For untargeted metabolite analysis, a CV cutoff of 20% and an average intensity value of 107 are used to filter the components table. These rigid CV and average intensity threshold values are used for this demonstration aim. CV cutoff values can be increased (for example, 30 %) while the average intensity values need to be decreased (for example, 105) to include more peaks. After manually checking peaks, components with good shapes are selected and searched for in the human metabolome database. The results are shown in Table 2. Table 2A lists the results from data collected in positive mode, while Table 2B shows the results from negative mode. Some of the metabolites identified here overlap with the metabolites in the targeted list, such as glutathione, proline and so on, but meanwhile, additional metabolites absent from the targeted list are explored, such as methyglyoxal, which can be derived from glycolysis, and 1-palmitoyl-2-oleoyl-sn-glycero-3-phosphocholine, which is detected in positive with a retention time of 3.2 min, which is a reasonable retention for phospholipids on an amide column. A protocol on untargeted metabolite database searching has been previously reported20.

Table 2.

List of untargeted metabolites detected in HCT 8 cells (2.78×105 cells equivalence).

Table 2A and 2B include components extraction information: retention time, m/z and mass error, and meanwhile, database search results: Chemspider ID (CSID), name, formula, and so on. Here the samples analyzed are equal to metabolites extracted from 2.78×105 cells, and the intensity threshold is 1×107 to avoid tedious result for demonstration aim.

CSID Name Formula Monoisoto pic Mass Search Mass DeltaPPM R.T. (min)
857 Methylglyoxal C3H4O2 72.02100 72.02108 1.03 7.70
1057 Sarcosine C3H7NO2 89.04768 89.04747 2.33 8.19
5735 alanine C3H7NO2 89.04768 89.04747 2.33 8.19
234 beta-Alanine C3H7NO2 89.04800 89.04747 5.93 8.19
55423 R-lactic acid C3H6O3 90.03169 90.03143 2.91 5.12
61460 Hydroxypropionic acid C3H6O3 90.03169 90.03143 2.91 5.12
96860 L-(+)-lactic acid C3H6O3 90.03169 90.03143 2.91 5.12
592 Lactic acid C3H6O3 90.03200 90.03143 6.29 5.12
650 Dihydroxyacetone C3H6O3 90.03200 90.03143 6.29 5.12
731 Glyceraldehyde C3H6O3 90.03200 90.03143 6.29 5.12
1086 Sulfuric acid H2O4S 97.96738 97.96683 5.63 8.13
128566 Proline C5H9NO2 115.06333 115.06302 2.67 7.78
1078 Succinic acid C4H6O4 118.02661 118.02630 2.66 7.72
466979 Erythrono-1,4-lactone C4H6O4 118.02661 118.02630 2.66 7.72
4483398 D-Erythronic g-lactone C4H6O4 118.02661 118.02630 2.66 7.72
473 Methylmalonic acid C4H6O4 118.02700 118.02630 5.96 7.72
8527138 (3S,4R)-3,4-Dihydroxydihydrofuran-2(3H)-one C4H6O4 118.02700 118.02630 5.96 7.72
140384 2-ketocaproic acid C6H10O3 130.06299 130.06270 2.24 2.35
164251 Methyloxovaleric acid C6H10O3 130.06299 130.06270 2.24 2.35
388419 (3S)-3-Methyl-2-oxopentanoic acid C6H10O3 130.06299 130.06270 2.24 2.35
15642233 Ketoleucine C6H10O3 130.06299 130.06270 2.24 2.35
46 a-Oxo-b-methylvaleric acid C6H10O3 130.06300 130.06270 2.36 2.35
69 Alpha-ketoisocaproic acid C6H10O3 130.06300 130.06270 2.36 2.35
134 Aminolevulinic acid C5H9NO3 131.05800 131.05795 0.36 8.19
9312313 3-Hydroxy-L-proline C5H9NO3 131.05800 131.05795 0.36 8.19
5605 HYDROXYPROLINE C5H9NO3 131.05824 131.05795 2.23 8.19
7068 N-Acetylalanin C5H9NO3 131.05824 131.05795 2.23 8.19
79449 Ac-Ala-OH C5H9NO3 131.05824 131.05795 2.23 8.19
89122 Ethylformylglycine C5H9NO3 131.05824 131.05795 2.23 8.19
167744 l-Glutamic-gamma-semialdehyde C5H9NO3 131.05824 131.05795 2.23 8.19
388519 5-Amino-2-oxopentanoic acid C5H9NO3 131.05824 131.05795 2.23 8.19
5880 L-(+)-Leucine C6H13NO2 131.09464 131.09419 3.39 7.09
6067 L-(+)-Isoleucine C6H13NO2 131.09464 131.09419 3.39 7.09
19964 L-Norleucine C6H13NO2 131.09464 131.09419 3.39 7.09
388796 beta-Leucine C6H13NO2 131.09464 131.09419 3.39 7.09
548 Aminocaproic acid C6H13NO2 131.09500 131.09419 6.18 7.09
109 Ureidopropionic acid C4H8N2O3 132.05299 132.05321 1.60 8.28
6031 L-(-)-Asparagine C4H8N2O3 132.05350 132.05321 2.21 8.28
6026 L-Ornithine C5H12N2O2 132.08987 132.08961 1.98 10.36
64236 D-Ornithine C5H12N2O2 132.08987 132.08961 1.98 10.36
193317 L-( )-Malic acid C4H6O5 134.02153 134.02130 1.72 7.95
510 (±)-Malic Acid C4H6O5 134.02200 134.02130 5.25 7.95
133224 threonic acid C4H8O5 136.03717 136.03688 2.14 7.70
388628 2,3,4-Trihydroxybutanoic acid C4H8O5 136.03717 136.03688 2.14 7.70
2061231 DL-erythronic acid C4H8O5 136.03717 136.03688 2.14 7.70
21436 N-Methyl-D-aspartic acid C5H9NO4 147.05316 147.05299 1.16 8.05
21814 D-(-)-Glutamic acid C5H9NO4 147.05316 147.05299 1.16 8.05
30572 L-(+)-Glutamic acid C5H9NO4 147.05316 147.05299 1.16 8.05
58744 N-Acetyl-L-serine C5H9NO4 147.05316 147.05299 1.16 8.05
6038 Histidine C6H9N3O2 155.06947 155.06930 1.09 8.36
199 Allantoin C4H6N4O3 158.04401 158.04387 0.88 4.76
6082 L-(+)-Arginine C6H14N4O2 174.11168 174.11154 0.80 10.76
64224 D-Arg C6H14N4O2 174.11168 174.11154 0.80 10.76
58576 N-Acetyl-L-Aspartic acid C6H9NO5 175.04807 175.04803 0.18 7.87
996 Pyrophosphoric Acid H4O7P2 177.94299 177.94331 1.79 8.42
5589 Glucose C6H12O6 180.06339 180.06346 0.41 7.53
17893 Mannose C6H12O6 180.06339 180.06346 0.41 7.53
58238 .beta.-D-Glucopyranose C6H12O6 180.06339 180.06346 0.41 7.53
71358 .alpha.-D-Glucopyranose C6H12O6 180.06339 180.06346 0.41 7.53
134838 3-deoxy-arabino-hexonic acid C6H12O6 180.06339 180.06346 0.41 7.53
388332 L-Sorbopyranose C6H12O6 180.06339 180.06346 0.41 7.53
388476 beta-D-galactopyranose C6H12O6 180.06339 180.06346 0.41 7.53
388480 .alpha.-D-Galactopyranose C6H12O6 180.06339 180.06346 0.41 7.53
388775 beta-D-Fructofuranose C6H12O6 180.06339 180.06346 0.41 7.53
10239179 Inositol C6H12O6 180.06339 180.06346 0.41 7.53
16736992 Cis-inositol C6H12O6 180.06339 180.06346 0.41 7.53
17216070 allose C6H12O6 180.06339 180.06346 0.41 7.53
17216093 L-Sorbose C6H12O6 180.06339 180.06346 0.41 7.53
201 hexopyranose C6H12O6 180.06300 180.06346 2.53 7.53
868 1,2,3,4,5,6-cyclohexanhexol C6H12O6 180.06300 180.06346 2.53 7.53
2068 Theophylline C7H8N4O2 180.06473 180.06346 7.04 7.53
4525 1,7-dimethyl-Xanthine C7H8N4O2 180.06473 180.06346 7.04 7.53
5236 Theobromine C7H8N4O2 180.06473 180.06346 7.04 7.53
1161 isocitric acid C6H8O7 192.02701 192.02704 0.15 8.18
305 Citric acid C6H8O7 192.02699 192.02704 0.23 8.18
963 pantothenic acid C9H17NO5 219.11067 219.11049 0.84 6.72
6361 D-pantothenic acid C9H17NO5 219.11067 219.11049 0.84 6.72
960 palmitic acid C16H32O2 256.23999 256.24000 0.05 1.73
111188 Glutathione C10H17N3O6S 307.08380 307.08339 1.35 8.03
388337 N-Acetylneuraminic acid C11H19NO9 309.10599 309.10585 0.45 7.43
392681 N-Acetyl-alpha-neuraminic acid C11H19NO9 309.10599 309.10585 0.45 7.43
392810 Sialic acid Neu5Ac C11H19NO9 309.10599 309.10585 0.45 7.43

Discussion

The most critical steps for successful metabolite profiling in cells using this protocol are: 1) controlling the growth medium and careful extraction of the cells; 2) adjusting the LC method based on MS method setup to ensure there are enough (usually at least 10) data points across a peak for quantitation; 3) doing a low mass calibration before running samples; 4) injecting no more than 5 μl to avoid retention time shifting and peak broadening caused by water; and 5) preparing and running samples for comparison in the same batch to minimize batch effects.

The standards (Table 1) chosen here for low mass range calibration are interchangeable. Any known compound with an m/z that falls into the mass scan range, is well behaved in a H-ESI source, and is soluble in water, methanol or acetonitrile are reasonable candidates for calibration standards. It is highly recommended to store all calibration solutions at 4 °C to stabilize caffeine and also to minimize the evaporation of methanol or acetonitrile in the calibration solution so that the calibration performance can be more reproducible. Compared to a regular mass range, m/z from 150 to 2000, low mass range calibration needs to be done more frequently, at least once every two days.

This workflow, from extraction solvent, reconstitution solvent, LC mobile phase, to low mass range calibration and MS scan range has been optimized to measure polar metabolites. This includes amino acids, acetyl amino acids, glycolysis pathway intermediates, nucleosides, TCA cycle intermediates, some one-carbon metabolism pathway intermediates and so on. However, modifications of this protocol for other classes of metabolites, such as Coenzyme A (CoA) species, folates, phospholipids are possible. For example, CoAs are more stable in acidic conditions, so the addition of an acid to 80% methanol/water will be helpful to improve CoA sensitivity. Also, CoAs and lipids tend to have much large molecular weights, thus the m/z scan range needs to be adjusted from 60–900 to the proper range which will cover those metabolites.

Even though some of the untargeted component database search results overlap with the targeted list, it is still of importance to build this targeted list based on the research priority. Since the metabolites in the targeted list are lower than the average intensity threshold, information on these metabolites will be removed during processing. The targeted list includes the retention time information, which gives us higher confidence for metabolite identification and quantitation. One further advantage with the QE-MS setup is that tandem mass spectrometry can allow for further identification of metabolites.

One issue associated with this workflow is that the H-ESI needle insert is sensitive to the salt content of the samples, as the sensitivity will be greatly compromised if there are high amounts of non-volatile salts. Therefore, minimizing salt content from samples, and routine cleaning of the column and the H-ESI needle insert will be helpful to ensure good quality data and to increase the column’s lifetime.

In summary, this protocol employs LC-QE-MS to successfully analyze polar metabolites from cultured cells, with minimal sample preparation steps and rapid data acquisition. Small modifications in sample preparation can be carried out to obtain data from other biological sources such as serum and tissue. For example, since pure methanol can be added to liquid serum added to make a final methanol concentration to 80% for polar metabolites extraction. For tissue samples, rigorous stirring and mixing is required to achieve better extraction efficiency. Usually 10 μl serum or 1 mg tissue is sufficient for metabolites analysis. The raw data can be analyzed both in targeted mode, if there are known metabolites in the samples, and in an un-targeted way followed by HRMS database searching. HRMS based metabolomics is still in its early stages. For future advances, the experimental techniques can be further optimized, additional metabolite HRMS information and MS/MS fragmentation patterns will be helpful and relevant algorithms, such as peak alignment, peak integration, isotope clustering and so on, can improve the efficiency and accuracy of data processing. Ultimately, however, many of the questions that our lab addresses in metabolism are limited by careful interpretation of data after processing. With these large-scale metabolomics techniques, we are often limited by our interpretation of the data and evaluation of hypotheses generated. Therefore, all metabolomics experiments need be formulated around specific questions.

Supplementary Material

Supplementary
supplementary 2

Acknowledgments

The authors would like to acknowledge Detlef Schumann, Jennifer Sutton (Thermo Fisher Scientific) and Nathaniel Snyder (University of Pennsylvania) for valuable discussions on mass calibration and data processing. Research reported in this publication was supported by the National Cancer Institute of the National Institutes of Health under Award Number R00CA168997. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Footnotes

A complete version of this article that includes the video component is available at http://dx.doi.org/10.3791/51358.

Disclosures: The authors declare no conflicts of interest.

References

  • 1.Fiehn O. Combining genomics, metabolome analysis, and biochemical modelling to understand metabolic networks. Comparative and Functional Genomics. 2001;2(3):155–168. doi: 10.1002/cfg.82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Mulleder M, et al. A prototrophic deletion mutant collection for yeast metabolomics and systems biology. Nature Biotechnology. 2012;30(12):1176–1178. doi: 10.1038/nbt.2442. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Patel VR, Eckel-Mahan K, Sassone-Corsi P, Baldi P. CircadiOmics: Integrating circadian genomics, transcriptomics, proteomics and metabolomics. Nature Methods. 2012;9(8):772–773. doi: 10.1038/nmeth.2111. [DOI] [PubMed] [Google Scholar]
  • 4.Ideker T, et al. Integrated genomic and proteomic analyses of a systematically perturbed metabolic network. Science. 2001;292(5518):929–934. doi: 10.1126/science.292.5518.929. [DOI] [PubMed] [Google Scholar]
  • 5.Jonsson P, et al. A strategy for identifying differences in large series of metabolomic samples analyzed by GC/MS. Analytical Chemistry. 2004;76(6):1738–1745. doi: 10.1021/ac0352427. [DOI] [PubMed] [Google Scholar]
  • 6.Jonsson P, et al. High-throughput data analysis for detecting and identifying differences between samples in GC/MS-based metabolomic analyses. Analytical Chemistry. 2005;77(17):5635–5642. doi: 10.1021/ac050601e. [DOI] [PubMed] [Google Scholar]
  • 7.Weljie AM, Newton J, Mercier P, Carlson E, Slupsky CM. Targeted profiling: Quantitative analysis of 1H NMR metabolomics data. Analytical Chemistry. 2006;78(13):4430–4442. doi: 10.1021/ac060209g. [DOI] [PubMed] [Google Scholar]
  • 8.Wiklund S, et al. Visualization of GC/TOF-MS-based metabolomics data for identification of biochemically interesting compounds using OPLS class models. Analytical Chemistry. 2008;80(1):115–122. doi: 10.1021/ac0713510. [DOI] [PubMed] [Google Scholar]
  • 9.Xia J, Bjorndahl TC, Tang P, Wishart DS. MetaboMiner--semi-automated identification of metabolites from 2D NMR spectra of complex biofluids. BMC Bioinformatics. 2008;9(507) doi: 10.1186/1471-2105-9-507. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Lu D, et al. 13C NMR isotopomer analysis reveals a connection between pyruvate cycling and glucose-stimulated insulin secretion (GSIS) Proceedings of the National Academy of Sciences of the United States of America. 2002;99(5):2708–2713. doi: 10.1073/pnas.052005699. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Shen J, et al. Determination of the rate of the glutamate/glutamine cycle in the human brain by in vivo 13C NMR. Proceedings of the National Academy of Sciences of the United States of America. 1999;96(14):8235–8240. doi: 10.1073/pnas.96.14.8235. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Kitteringham NR, Jenkins RE, Lane CS, Elliott VL, Park BK. Multiple reaction monitoring for quantitative biomarker analysis in proteomics and metabolomics. Journal of Chromatography. B, Analytical Technologies in the Biomedical and Life Sciences. 2009;877(13):1229–1239. doi: 10.1016/j.jchromb.2008.11.013. [DOI] [PubMed] [Google Scholar]
  • 13.Locasale JW, et al. Metabolomics of human cerebrospinal fluid identifies signatures of malignant glioma. Molecular & Cellular Proteomics : MCP. 2012;11(6) doi: 10.1074/mcp.M111.014688. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Wolf-Yadlin A, Hautaniemi S, Lauffenburger DA, White FM. Multiple reaction monitoring for robust quantitative proteomic analysis of cellular signaling networks. Proceedings of the National Academy of Sciences of the United States of America. 2007;104(14):5860–5865. doi: 10.1073/pnas.0608638104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Yuan M, Breitkopf SB, Yang X, Asara JM. A positive/negative ion-switching, targeted mass spectrometry-based metabolomics platform for bodily fluids, cells, and fresh and fixed tissue. Nature Protocols. 2012;7(5):872–881. doi: 10.1038/nprot.2012.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ramanathan R, et al. It is time for a paradigm shift in drug discovery bioanalysis: From SRM to HRMS. Journal of Mass Spectrometry : JMS. 2011;46(6):595–601. doi: 10.1002/jms.1921. [DOI] [PubMed] [Google Scholar]
  • 17.Lu W, Clasquin MF, Melamud E, Amador-Noguez D, Caudy AA, Rabinowitz JD. Metabolomic analysis via reversed-phase ion-pairing liquid chromatography coupled to a stand alone orbitrap mass spectrometer. Analytical Chemistry. 2010;82(8):3212–3221. doi: 10.1021/ac902837x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Michalski A, et al. Ultra high resolution linear ion trap orbitrap mass spectrometer (orbitrap elite) facilitates top down LC MS/MS and versatile peptide fragmentation modes. Molecular & Cellular Proteomics : MCP. 2012;11(3) doi: 10.1074/mcp.O111.013698. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Michalski A, et al. Mass spectrometry-based proteomics using Q exactive, a high-performance benchtop quadrupole orbitrap mass spectrometer. Molecular & Cellular Proteomics : MCP. 2011;10(9) doi: 10.1074/mcp.M111.011015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Snyder NW, Khezam M, Mesaros CA, Worth A, Blair IA. Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry (UPLC-HRMS) J Vis Exp. 2013;75:e50433. doi: 10.3791/50433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Gika HG, Theodoridis GA, Wingate JE, Wilson ID. Within-day reproducibility of an HPLC-MS-based method for metabonomic analysis: Application to human urine. Journal of Proteome Research. 2007;6(8):3291–3303. doi: 10.1021/pr070183p. [DOI] [PubMed] [Google Scholar]
  • 22.Want EJ, et al. Global metabolic profiling procedures for urine using UPLC-MS. Nature Protocols. 2010;5(6):1005–1018. doi: 10.1038/nprot.2010.50. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary
supplementary 2

RESOURCES